CW-6200 industrial water chiller is designed for laser as its target application, but it is also used for other various applications, such as laboratory device, rotary evaporator, medical equipment, induction heater and many more.
CW-6200 industrial water chiller is designed for laser as its target application, but it is also used for other various applications, such as laboratory device, rotary evaporator, medical equipment, induction heater and many more.
This air cooled chiller delivers temperature stability of ±0.5℃ together with 5100W cooling capacity. And thanks to the powerful water pump, water circulation between the chiller and the heat generating process can be ongoing to take away the heat. The standard temperature range for CW-6200 water chiller is 5-35 degree Celsius.
The warranty period is 2 years.
Features
1. 5100W cooling capacity. R-410a refrigerant with low global warming potential;
Specification
Note:
1. The working current can be different under different working conditions; The above information is for reference only. Please subject to the actual delivered product;
2. Clean, pure, impurity free water should be used. The ideal one could be purified water, clean distilled water, deionised water, etc.;
3. Change out the water periodically (every 3 months is suggested or depending on the actual working environment).
4. Location of the chiller should be well ventilated environment. There must be at least 50cm from the obstacles to the air outlet which is on the top of the chiller and should leave at least 30cm between obstacles and the air inlets which are on the side casing of the chiller.
PRODUCTION INTRODUCTION
User-friendly temperature controller for easy operation
Equipped with caster wheels for easy mobility
Water inlet and outlet ports made from stainless steel to prevent potential corrosion or water leakage.
Easy-to-read water level check. Fill the tank until the water reaches the green area.
Cooling fan of famous brand installed.
Alarm description
CW-6200 industrial water chiller is designed with built-in alarm functions.
E1- ultrahigh room temperature
E2 - ultrahigh water temperature
E3 - ultralow water temperature
E4 - room temperature sensor failure
E5 - water temperature sensor failure
E6 - external alarm input
E7 - water flow alarm input
CHILLER APPLICATION
WAREHOUSE