Water ORP Transmitter Detection Sensor

Sales Water ORP Transmitter Detection Sensor

♦Adopt double high-impedance three-electrode system, with online one-key calibration, real-time temperature compensation, electrode loose alarm, electrode good or bad alarm during calibration, and power-off protection (the calibration result and preset data will not be lost due to shut down or power failure).

♦Over-current protection, over voltage protection, high measurement accuracy, fast response, long life and so on.

♦Small in size, light in weight, easy to install and maintain, and the standard industrial signal output (0-5V, 0-10V, 4-20MA, Modbus RTU485) can meet the limit of on-site connection to various real-time monitoring equipment.

Product Detail  

Water ORP Transmitter Detection Sensor

ORP transmitter adopt double high-impedance three-electrode system, with online one-key calibration, real-time temperature compensation, electrode loose alarm, electrode good or bad alarm during calibration, and power-off protection (the calibration result and preset data will not be lost due to shut down or power failure) , Over-current protection, over voltage protection, high measurement accuracy, fast response, long life and so on.

This product is suitable for: electroplating, yeast, food processing, sewage treatment, metallurgy, environmental protection and other occasions.

ltems Parameters
Measurement ORP range -1000~1000mV or-2000~2000mV
Measuring ORP Accuracy ±3mV@-1000~1000mV
±6mV@-2000~2000mV
ORP sensitivity 1mV
Power supply DC 12V-24V
Output 0-5V/0-10V/4-20mA/RS485
load resistance Voltage output: R load 2 10K Ω
Current output: R loads(Uvcc-310.02Q
Consumption <1W
Storage condition 10-50℃(-20~+80°℃ peak) 20-60%RH
Working environment -20℃~+80℃
dimension 65X46X28.5mm³

The linear relation between analog signals and the corresponding physical value:

0 to 5 v output:

(-1000~1000mV range):ORP=ORP output voltage*400-1000;

(-2000~2000mV range):ORP=ORP output voltage*800-2000;

0 to 10 v output:

(-1000~1000mV range):ORP=ORP output voltage*200-1000;

(-2000~2000mV range):ORP=ORP output voltage*400-2000;4-20 ma output:

(-1000~1000mV range):ORP=(ORP current-4)*125-1000;

(-2000~2000mV range):ORP=(ORP current-4)*250-1000;

 

Due to the ORP electrode in the measurement environment vulnerable to pollution cause electrode signal affected, in order to ensure the ORP sensor can normally use for a long time, usually need to make a correction on a regular basis through the calibration of transducer (once a month). Depending on the water samples of specific calibration cycle.

FAQ
Q2. Can I have a free sample for testing ?
A: Sample order is acceptable, but need you to pay for the product cost and freight .

Q3. What about the lead time?
A: Sample needs 4-5 working days after receiving your payment, mass production needs extra 4-5 working days.

 


Q4: Is it OK to print my logo on the products?
A: Yes, we support OEM/ODM service.
Leave A Message
If you are interested in our products and want to know more details,please leave a message here,we will reply you as soon as we can.
X

Home

Supplier

Leave a message

Leave a message

If you are interested in our products and want to know more details,please leave a message here,we will reply you as soon as we can.